The Ultra SlimPak II is an exciting new line
of isolating signal conditioners from Action Instruments with greater accuracy
and better stability than virtually any other signal conditioners on the market
today.
The WV128 is a Thermocouple Input Limit Alarm
with dual setpoints and two contact closure outputs. The input type and alarm
functions are field configurable, offering maximum flexibility. There are seven
thermocouple types (B, E, J, K, R, S and T) from which to choose. Alarm
setpoints are fully adjustable throughout the entire range.
The WV128-2000 contains two standard
non-latching SPDT relays. When power is removed, the relays revert to their
non-powered state. The WV128-2001 contains two latching relays. Whatever
position the relays are in when power is removed is where they will remain when
power is re-applied (a reset switch is provided to reset the relays after the
alarm is no longer true). Both models are configurable as a single or dual
setpoint alarm, with HI or LO trips. The WV128-2000 supports failsafe or
non-failsafe operation. In failsafe operation the relay coil is energized when
the process is below the HI setpoint or above the LO setpoint (opposite for
non-failsafe). In the failsafe mode, a power failure results in an alarm state
output.
Operation Each of the alarm setpoints has a
respective HI or LO deadband. Red LEDs will illuminate indicating an alarm
condition. The trip will reset only when the process falls below the HI
deadband or rises above the LO deadband. (The WV128-2001 requires pressing the
reset switch in order to reset the relays after being tripped.) The deadband
can be left at the minimum level (the factory default) or can be set to
whatever level is desired. |